Database Inconsistency Failure
Symptom
Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited. Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. Database inconsistency failure and a Debug text of Check And Repair Requested at Browser initialization. When the customer boots the system, the image monitor displays Database inconsistency failure: Continue and call service for repair. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited.
Error message
1200012 RTAC_C80_Boards Async error card2 ( IP3 ) c80 Error level0_user_error :id:1_param 1200012 err_c80_proc components RTAC_c80_Boards, debug text Async error_card2_c80_Error level0_user_error :id:1_param 1800018 Error_sib_ip_boards_links [dsp_automatic Naming ] Ip2(card1)_IP3(card2) link failure detected 120001 error_soft 1200012 RTAC_C80_Boards Async error card2 c80 Error level0_user_error :id:1_param 1800018 ERR SIB IP BOARDS LINKS IP2 (card 1) -IP3 (card 2) link failure detected
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.