ASSET Calibration Scan Fails Due to Coil.
Symptom
ASSET Calibration Scan Fail See Long Description The system failed only with this type of scans, everything else worked fine.
Error message
2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Sun Nov 20 12:49:19 2005 Error: 2254782 Host: HDMR Process: NSP File: SCP:UPM Line: 657 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Peak value detected = 2236 2254791 The Universal Power Monitor X detected a Pulse Width fault on the Narrow Band Body Amplifier Sun Nov 20 12:43:16 2005 Error: 2254791 Host: HDMR Process: NSP File: SCP:UPM Line: 1231 The Universal Power Monitor 1 detected a Pulse Width fault on the Narrow Band Body Amplifier. Detected Width = 130 uS. Max Width Limit = 0 uS. 2254782 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Mon Aug 14 18:50:31 2006 Error: 2254782 Host: MMCLXMR1 Process: NSP File: SCP:UPM Line: 657 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Body Amplifier. Peak value detected = 13444 W.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion 1.5T 8 Channel TR Knee Array 5114356-2 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.