SSM HVON and DRV FLT Rocker Switches Were In the "DIS" Position.Unable To Scan With Driver Module Errors.Set the SSM HVO…
SSM HVON and DRV FLT Rocker Switches Were In the "DIS" Position.Unable To Scan With Driver Module Errors.Set the SSM HVON and DRV FLT (MC,TR,DD) Rocker Switches to the "EN" Position wqr.Signa HDxT.225…
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252289 Driver Module is in Service Mode. A Cause is DD Fault Disable Switch. Error: 2252289 Host: MR3T Process: NSP File: SCP:DCB Line: 933 Driver Module is in Service Mode.^A Cause is DD Fault Disable Switch.-> 2249163 SSM fault. High voltage power supply low. 2249163 Host: mr01 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2370 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Gradient Echo Coil: 8HRBRAIN Scan: 26489/1 SSM fault.^A High voltage power supply low. 2249163 Host: mr01 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2370 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Gradient Echo Coil: 8HRBRAIN Scan: 26489/1 SSM fault. High voltage power supply low. 2252289 Driver Module is in Service Mode. A Cause is High Voltage Disable Switch. Error: 2252289 Host: MR3T Process: NSP File: SCP:DCB Line: 933 Driver Module is in Service Mode.^A Cause is High Voltage Disable Switch.->
Applies to
- Models
- Signa, Excite
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.