Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova5020438002002400015

X-ray unavailable due to error code = 2400015 followed by error code = 50204 followed by error code = 3800200

Symptom

X-ray unavailable due to error code = 2400015 followed by error code = 50204 followed by error code = 3800200 X-ray unavailable due to error code = 2400015 followed by error code = 50204 followed by error code = 3800200. Injector issue. Can be reproduced by power-on the MEd RAdV injector after start exam, and try X-rays.

Error message

error Code = 50204, error Text = Injector started at an unexpected moment , debug Text += pos-error=4A72, recovery Class = AR, major Function /= XRImChainPos System impact: X-ray unavailable. Displayed message : "Injector start pb. Stop X-ray, reset system. If pb persists call service"

error Code = 3800200, error Text = State machine at wrong state, debug Text /= got TO_START_PREPARE at: state OpModeConfigure, _configureStatus = 256, recovery Class = AR, major Function /= XRTechOpt System impact: X-ray unavailable. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service"

error Code = 2400015, error Text = Injector started at an unexpected moment , debug Text += Unexpected injection, please release X-ray switches, recovery Class = AR, major Function /= XRPeriDevCtrl System impact: X-ray unavailable. Displayed message : "Injector start pb. Stop X-ray, reset system. If pb persists call service"

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Injector Injector Injector
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.