GE Healthcare · MRExciteSigna22538222248759
Rf amp response timeout due to SSM.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2253822 The MGD could not send a packet to one of the MDS peripherals. 2253822 Proc:Diag The MGD could not send a packet to one of the MDS peripherals. Because of this, the diagnostic currently running was halted. Run MDS Link diagnostics. 2248759 RF Amplifier Response Timeout RF MDS peripheral to RF Amp/RF Interface link failure. 2248759 NSP SCP: RF amplifier response timeout. RF MDS peripheral - RF Amp/RF Interface link failure. Verify that the RF Amplifier or RF Interface is powered on and that the RF MDS peripheral is connected to the RF Amplifier or RF Interface.
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.