Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova12001041200014

X-ray not Available shown on the acquisition monitor at boot up.

Symptom

X-ray not Available shown on the acquisition monitor at boot up. Several issues that are related to keeping the detector from Booting. Detector boot issues will cause the Rtac not to boot Tube chiller

Error message

1200104, error Text = ERR_DET_POWER_SUPPLY, debug Text += [DemSupervisor::_PWSStatusDebugMessage] Statuses are bad : (PG1B : OK) (PG2A : OK) (PG2B : OK) (PG3 : BAD) (PG1A : OK)~ (Retry number: 1)~ Note: PG3 bad indicates that PS defect is related to the logic voltage used especially for fiber channel communication. Thus we may also get detector comm error 1200014 in the system errlog. 1200104, error Text = ERR_DET_POWER_SUPPLY, debug Text += [DemSupervisor::_PWSStatusDebugMessage] Statuses are bad : (PG1B : OK) (PG2A : OK) (PG2B : OK) (PG3 : BAD) (PG1A : OK)~ (Retry number: 1)~ Note: PG3 bad indicates that PS defect is related to the logic voltage used especially for fiber channel communication. Thus we may also get detector comm error 1200014 in the system errlog.

error Code = 1200104, error Text = ERR_DET_POWER_SUPPLY, debug Text += [DemSupervisor::_PWSStatusDebugMessage] Statuses are bad : (PG1B : OK) (PG2A : OK) (PG2B : BAD) (PG3 : OK) (PG1A : OK)~~, recovery Class = IH, major Function /= XRImDet System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service". error Code = 1200104, error Text = ERR_DET_POWER_SUPPLY, 

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Detector and Detector Power Supply Detector PS 2302361-2 for 21 cm detector, 2302361-3 (alt # 2300475-2) for 31/41 detector Detector PS 2302361-2 for 21 cm detector, 2302361-3 (alt # 2300475-2) for 31/41 detector
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.