Shutdown Software and Power Off the PDU.
Symptom
Can Not Scan
Error message
2248356 SSI Watchdog Timeout. An SSI Was Not Detected in the Timeout Period. See Long Description Wed Jul 7 09:33:17 2004 Error: 2248356 Host: GEMRI_1Process: NSP File: SCP:sct_wdog_exp Line: 852 SSI Watchdog Timeout. An SSI was not detected in the timeout period. The PSD or AGP might not be running. It may be necessary to perform a TPS reset. If TPS Reset doesn't help then AC 60Hz clock may be missing. Extended Error: Check the 60Hz clock signal from PDM to STIF Board in CAM chassis. Run CAM Data Acquisition SRF Diagnostics.
Applies to
- Models
- Signa, Excite, Discovery, Optima
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.