Pular para o conteúdo principal
GE Healthcare · MRExcite

PDU Main Breaker Trips, Will Not Reset due to Failed TVSS Module.

Symptom

Phoenix PDU Main Breaker Trips, Will Not Reset

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2185383-11 MOV MODULE Also known as "TVSS". When FE's attempt to order the TVSS module listed in HFO3 Renewal Parts: 2228875-7, FEMC automatically "changes up" to p/n 2185383-11 MOV MODULE instead, because 2228875-7 cannot be obtained anymore. Phoenix is out of business. I checked with Margie Wiza of MR Eng and this is as intended. This is to fix PQR # 13109044.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues