GE Healthcare · MRExciteSigna2254782
PM EPIWP is reusing the TG value found during the previous run of LVshim.
Symptom
PM Head EPIWP prematurely ends and Power Monitor Peak Power Exceeded Faults are visible in Error Log Effectivity: All Signa 1.5T and 3T EXCITE HD (12.0) MR Systems. Problem Description: After running LVshim the Head PM EPIWP (EPI White Pixel) test will complete quickly and not generate data. Body EPIWP will complete normally and generate data. A review of the error log will show that a Head Power Monitor Peak Power Exceeded Fault ocurred and caused the Head portion of the test to end prematurely. (SMILe Nov. 10, 2005)
Error message
2254782 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 1 df15hdmr NSP SCP:UPM 657 Exception Class: Unknown Severity: Unknown Function: RF: RF Power Monitor PSD: Spin Echo Coil: HEAD Scan: 7975/1 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Ban d Head Amplifier. Check patient weight entry. Peak value detected = 657 W. Max Peak Limit = 626 W- 2254782 The Universal Power Monitor 1 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. 2254782 NSP SCP:UPM 657 The Universal Power Monitor 2 has detected Peak Power Exceeded on the Narrow Band Head Amplifier. Peak value detected = 429 W. Max Peak Limit = 426 W- Possible cause(s): ASC RF Detector, ASC RF Processor, RF Amp, Software .
Applies to
- Models
- Excite, Signa
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.