GE Healthcare · X-ray VascularInnova1050222
TSSC Failure.
182243 TSSC Failure.
Symptom
TSSC Failure. Call Service is displayed on the Innova review monitors.
Error message
1050222 Master TSSC: Collimator Satellite error Error code 1050222 Master TSSC: Collimator Satellite error. Recommended Action: Check for loose set screw on joystick, see long for TSG. Check that TSSC is well plugged on 'Status Control M' connector on table foot. Run the Master TSSC diagnostics. Check that the satellite is well connected inside the TSSC box. can be caused by a loose set screw If still problem, change the collimator satellite. 1050222: Root Cause: Master TSSC: Collimator Satellite error This problem appears on the Master TSSC Root cause Probability Master TSSC Collimator Satellite High Master TSSC Low Recommended Action: Check that TSSC is well plugged on 'Status Control M' connector on table foot. Check for loose set screw on joystick Run the Master TSSC diagnostics. Check that the satellite is well connected inside the TSSC box. If still problem, change the collimator satellite.
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.