182243 TSSC Failure.

Symptom

TSSC Failure. Call Service is displayed on the Innova review monitors.

Error message

1050222 Master TSSC: Collimator Satellite error Error code 1050222 Master TSSC: Collimator Satellite error. Recommended Action: Check for loose set screw on joystick, see long for TSG. Check that TSSC is well plugged on 'Status Control M' connector on table foot. Run the Master TSSC diagnostics. Check that the satellite is well connected inside the TSSC box. can be caused by a loose set screw If still problem, change the collimator satellite. 1050222: Root Cause: Master TSSC: Collimator Satellite error This problem appears on the Master TSSC Root cause Probability Master TSSC Collimator Satellite High Master TSSC Low Recommended Action: Check that TSSC is well plugged on 'Status Control M' connector on table foot. Check for loose set screw on joystick Run the Master TSSC diagnostics. Check that the satellite is well connected inside the TSSC box. If still problem, change the collimator satellite.

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100