Image Quality Degraded; Tube inspection requested displayed during fluoro or Record, may hear loud popping noise coming…

Image Quality Degraded; Tube inspection requested displayed during fluoro or Record, may hear loud popping noise coming from tube

Symptom

JEDI: Tube Spit or kV instability, steady state error Related JEDI: Tube Spit or kV instability, steady state error Related. X-rays acquisition can be aborted if too many tube spits are occuring. JEDI: Spit or kV instability, steady state error Related. X-rays acquisition can be aborted if too many tube spits are occuring.

Image Quality Degraded; Tube inspection requested displayed during fluoro or Record, may hear loud popping noise coming from tube Image Quality Degraded; Tube inspection Requested displayed during fluoro or Record, may hear loud popping noise coming from tube Image Quality Degraded. Tube inspection Requested displayed during fluoro or Record , may hear loud popping noise coming from tube.

Error message

3602016 Degraded image quality, tube inspection requested WR XRTechMgr acquisition Component(s)_Tube 1 0 [XRTechMgr::notifyGeneratorFinalExposureParameters] numberOfTubeSpits=7 above the maxNumberOfTubeSpits=2 after 2 frames XRTechMgr.cpp 1958 3602016 "Degraded image quality, tube inspection requested" WR XRTechMgr acquisition Component(s)_Tube 1 0 [XRTechMgr::notifyGeneratorFinalExposureParameters] numberOfTubeSpits=7 above the maxNumberOfTubeSpits=2 after 2 frames XRTechMgr.cpp 1958

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Ulysses MX160 X-Ray Tube Ulysses MX160 X-Ray Tube Part number 2216500 (new) or D2802A (reload) Ulysses MX160 X-Ray Tube Part number 2216500 (new) or D2802A (reload)