Site Air Conditioning failure. 3T RF Amplifier Fails with PA Air Too Hot Fault 47.
Symptom
3T RF Amplifier Fails with PA Air Too Hot Fault 47
Error message
2238930 3T RF Amplifier PA Air Too Hot (fault 47). See Long Description Error 2238930 3T RF Amplifier PA Air Too Hot (fault 47). Possible Causes: obstructed air flow, PSD exceeding safe operating range. Cool 30 minutes, clear air intake and exhaust port, adjust PSD parameters. If the fault persists, replace the RF amplifier.
Applies to
- Product
- 3T Eclipse For E2 software
- Software
- VH3 3T VH3
- Area
- MR
Full cause & step-by-step resolution