Pular para o conteúdo principal
GE Healthcare · MRExciteDiscoveryOptima2249320108193483022493212248602224930720000

Cannot Boot Applications with Host Monitor Timeout Error.

Symptom

Cannot Boot Applications with Host Monitor Timeout Error

Error message

2249320 Communication problem between the APS, AGP, and SCP boards. Timeout: apsProbeComplete not received. 1081934830 6 1 Wed Apr 14 09:27:10 2004 2249320 2 CARD diagCoord systemResourceApi.c 525 Task ID: 0x3f1fdf8 Time: 3249 Object: AGP Communication problem between the APS, AGP, and SCP boards. Timeout: apsProbeComplete not received.->

2249321 Communication problem between the APS, AGP, and SCP boards. Timeout: apsConvertComplete not received. 1081934830 8 1 Wed Apr 14 09:27:10 2004 2249321 2 CARD diagCoord systemResourceApi.c 570 Task ID: 0x3f1fdf8 Time: 5254 Object: AGP Communication problem between the APS, AGP, and SCP boards. Timeout: apsConvertComplete not received.->

2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed.

2249307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Error seen with MGD Diags: 1081934830 0 1 Wed Apr 14 09:27:10 2004 2249307 2 CARD diagCoord pearlMap.c 1806 Task ID: 0x0 Time: 0 Object: SCP Communication problem between the APS, AGP, and SCP board

Applies to

Models
Excite, Discovery, Optima
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.