GE Healthcare · MRExciteDiscoveryOptima2249320108193483022493212248602224930720000
Cannot Boot Applications with Host Monitor Timeout Error.
Symptom
Cannot Boot Applications with Host Monitor Timeout Error
Error message
2249320 Communication problem between the APS, AGP, and SCP boards. Timeout: apsProbeComplete not received. 1081934830 6 1 Wed Apr 14 09:27:10 2004 2249320 2 CARD diagCoord systemResourceApi.c 525 Task ID: 0x3f1fdf8 Time: 3249 Object: AGP Communication problem between the APS, AGP, and SCP boards. Timeout: apsProbeComplete not received.-> 2249321 Communication problem between the APS, AGP, and SCP boards. Timeout: apsConvertComplete not received. 1081934830 8 1 Wed Apr 14 09:27:10 2004 2249321 2 CARD diagCoord systemResourceApi.c 570 Task ID: 0x3f1fdf8 Time: 5254 Object: AGP Communication problem between the APS, AGP, and SCP boards. Timeout: apsConvertComplete not received.-> 2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2249307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Error seen with MGD Diags: 1081934830 0 1 Wed Apr 14 09:27:10 2004 2249307 2 CARD diagCoord pearlMap.c 1806 Task ID: 0x0 Time: 0 Object: SCP Communication problem between the APS, AGP, and SCP board
Applies to
- Models
- Excite, Discovery, Optima
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.