GE Healthcare · MRSignaExcite2252245
Intermittent CAN link errors due to Driver Module
Symptom
Intermittent CAN Link Errors
Error message
2252245 Network Management Recovery detected on the CAN network. The node reset and recovered. Recovered Node is Master Gradient. 2252245 IMIMDLMR2 NSPSCP:CanLinkHWO 812Exception Class: Unknown Severity: UnknownFunction: Peripherals: CommunicationsPSD: Unknown Coil: Unknown Scan: -1/-1Network Management Recovery detected on the CAN network.The node reset and recovered.Recovered Node is Master Gradient.
Applies to
- Models
- Signa, Excite
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.