GE Healthcare · MRExcite2244720224715951174

Defective Master Exciter2 board.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2244720 DSP2 on the GP did not respond to the last interrupt request (Handshake Error). Please perform a TPS Reset and try again See Long Description 2244720 BROCKGEMR78 NSP SCP:Grad 2339 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 DSP2 on the GP did not respond to the last interrupt request (Handshake Error). Please perform a TPS Reset and try again.->

2247159 Sine Wave Table on RIF DIF1 Board Did Not Load Correctly See Long Description Fri Jan 25 09:23:14 2008 Error: 2247159 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 412 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51174 Object: APS Test Task The Sine Wave Table on the DIF board in RRF chassis (front, slot 10) did not load correctly. This will prevent proper Exciter output. Try TPS_reset. Run RRF diagnostics. Reported from Irf::reset_rrf (0) (0) (150)->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.