GE Healthcare · MRExciteSigna2244570224485122445772244578
2244851 Current distortion fault due to HFD-S Lite Gradient Control Board.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244570 The gradient amplifier failed to go to Ready when commanded. 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.-> 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded. 2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault. the same error as 2244577 on 11X, G3, openMR4 and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5. See Long Description 2244851 error code on Post 14M5 and 12.0 M5, is the same error as 2244577 on 11X, G3, openMR4and 2244578 on 12X, E2, 14M4 (HDx, HDe), and openMR5.2244851 The GP is reporting a Gradient Amplifier %c Axis Current Distortion Fault Example error message (could be any gradient axis, Y-axis is shown in this example): Error: 2244851 Host: lx-mr2 Process: NSP File: SCP:Grad Line: 2851 The GP is reporting a Gradient Amplifier Y Axis Current Distortion Fault.-> Extended error message This is a gen
Applies to
- Models
- Excite, Signa
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Component
- SecondOpinion Control Board Assembly HFD-S-LITE 5138035-2
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.