Unable to Scan with SGA Power Supply UnderVoltage Fault due to SGA-PS-LITE.

Symptom

Unable to Scan with SGA Power Supply UnderVoltage Fault See Long Description All HFA-S Lite gradient amplifiers are showing UV or undervoltage on power up, and there are undervoltage errors in the log (see below for error messages) and the system won't scan. Also HFA-S Lite Power Supply (PS-Lite2) is showing UV for all three axes. Cycling complete system power doesn't help. Fri Jan 25 15:04:37 2008 2252245 IMIMDLMR2 NSP SCP:CanLinkHWO 812 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 Network Management Recovery detected on the CAN network. The node reset and recovered. Recovered Node is Master Gradient. Fri Jan 25 15:04:43 2008 2244857 IMIMDLMR2 NSP SCP:Grad 746 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The GP is reporting SGA Power Supply Power Off Fault.-> Fri Jan 25 15:04:51 2008 2244862 IMIMDLMR2 NSP SCP:Grad 3125 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The GP is reporting a SGA Power Supply Under Voltage Fault.-> Fri Jan 25 15:04:52 2008 2244607 IMIMDLMR2 NSP SCP:Grad 3142 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The GP reported multiple SGA Power Supply UnderVoltage Faults.->

Error message

2244862 The GP is reporting a SGA Power Supply Under Voltage Fault. Same as 2244592 and 2244593 2244862 error code on Post 14M5 is the same error as 2244592 on 11X, G3 and 2244593 on 12X, E2, 14M4 (HDx, HDe). 2244862 The GP is reporting a SGA Power Supply Under Voltage Fault. 2244862 error code on Post 14M5 is the same error as 2244592 on 11X, G3 and 2244593 on 12X, E2, 14M4 (HDx, HDe).2244862 The GP is reporting a SGA Power Supply Under Voltage Fault.

2224666 Auto Prescan failed, Master Gradient is not ready

2244607 The GP reported mutiple SGA Power Supply UnderVoltage Faults. same as 2244605 11X, G3, and OpenMR4. 2244607 The GP reported mutiple SGA Power Supply UnderVoltage Faults. 2244607 on 12X, 14X(HDx, HDe), E2, OpenMR5 sw is the same as 2244605 11X, G3, and OpenMR4. 2244607 The GP reported mutiple SGA Power Supply UnderVoltage Faults.

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
5138099-2 PS-Lite2 Gradient Amplifier with FRU packing 5138099-2 PS-Lite2 Gradient Amplifier with FRU packing
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.