Multicoil Bias Faults Due to Coil.

Symptom

Unable to Scan With Multi Coil Bias Faults Unable to scan with Multi Coil Bias Faults

Error message

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2224666 Auto Prescan failed, Auto Prescan failed, possible data acquisition failure. 2224666 mmp-mr01 psc PscMainControl.cc 1311 Auto Prescan failed, possible data acquisition failure.

2252301 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCDx Channel x of y, counting from 1 Value x mA Limit y mA See Long Description for Extended Error 2252301 MRI_GB_OW0 NSP SCP:DCB 816 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 (or MCD2) Channel 1 of 16, counting from 1 Value: 2 mA Limit: 298 mA Mask of failed channels: 0x1 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. An open circuit fault occurs when the actual value is less than the pre-defined limit. Check system cables, connectors, coils and configurations. Re-seat all coil connectors. The mask of all failed channels indicates additional channels wi

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2293668-2 1.5T NEUROVASCULAR ARRAY W/CABLE
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.