GE Healthcare · X-ray VascularInnova1050222

INX100, Collimator will move in but cannot move out or Collimator is not working properly, blades close and drive into t…

INX100, Collimator will move in but cannot move out or Collimator is not working properly, blades close and drive into the FOV without being requested to move by the operator

Symptom

INX100, Collimator will move in but cannot move out or Collimator is not working properly, blades close and drive into the FOV without being requested to move by the operator INX100, Collimator will move in but cannot move out or Collimator is not working properly, blades close and drive into the FOV without being requested to move by the operator. Collimator reset not working on TSSC.

Error message

1050222 Master TSSC: Collimator Satellite error Error code 1050222 Master TSSC: Collimator Satellite error. Recommended Action: Check for loose set screw on joystick, see long for TSG. Check that TSSC is well plugged on 'Status Control M' connector on table foot. Run the Master TSSC diagnostics. Check that the satellite is well connected inside the TSSC box. can be caused by a loose set screw If still problem, change the collimator satellite. 1050222: Root Cause: Master TSSC: Collimator Satellite error This problem appears on the Master TSSC Root cause Probability Master TSSC Collimator Satellite High Master TSSC Low Recommended Action: Check that TSSC is well plugged on 'Status Control M' connector on table foot. Check for loose set screw on joystick Run the Master TSSC diagnostics. Check that the satellite is well connected inside the TSSC box. If still problem, change the collimator satellite.

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Innova Collimator Innova collimator Innova collimator
Area
X-ray Vascular

Full cause & step-by-step resolution