Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova1200113120010812001041200014

Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor.

Symptom

Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Image Monitor. Xray not Available - Reset System; If Problem Persists Call Service is displayed on the Monitor.

Error message

error Code = 1200113, error Text = ERR_DET_POWER_SUPPLY_BOOT_WARNING, debug Text += [DemSupervisor::_PWSStatusDebugMessage] Statuses are bad : (PG1B : BAD) (PG2A : BAD) (PG2B : BAD) (PG3 : BAD) (PG1A : BAD)~ followed by error Code = 1200108 System impact: System boot failure. Displayed message : "Xray not available - Reset system; if problem persists, call Service". 1200108: Root Cause: Component(s) - Frontal Detector Power Supply This error is due to bad status sent by Frontal Detector Power Supply to Frontal RTAC Gitane. Recommended Action: Run RTAC subsystem//Front Detector Power Supply diagnostics 1200113: Root Cause: Component(s) - Frontal Detector Power Supply Warning This error is due to an unsucessful Detector Power supply boot attempt for frontal plane. Recommended Action: If the error is NOT repeated and NOT followed by a DetPS Boot Failure error, no Service recommended action

1200104, error Text = ERR_DET_POWER_SUPPLY, debug Text += [DemSupervisor::_PWSStatusDebugMessage] Statuses are bad : (PG1B : OK) (PG2A : OK) (PG2B : OK) (PG3 : BAD) (PG1A : OK)~ (Retry number: 1)~ Note: PG3 bad indicates that PS defect is related to the logic voltage used especially for fiber chann

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues