Pular para o conteúdo principal
GE Healthcare · MRExciteDiscoveryOptima224942222279152218527

CWA Unable to Scan with Error No. 2227915 The RDBM Read Header function returned the following error 6 The IT link is po…

CWA Unable to Scan with Error No. 2227915 The RDBM Read Header function returned the following error 6 The IT link is possibly disconnected or the IT Boards are not communicating.

Symptom

Unable to Scan with Error No. 2227915 The RDBM Read Header function returned the following error 6 The IT link is possibly disconnected or the IT Boards are not communicating See Long Description The system gives the following errors immediately once autoprescan is passed and scan begins. This happens on all coils once this error occurs. 2227915 "The RDBM Read Header function returned the following error :6 The IT link is possibly disconnected or the IT Boards are not communicating." 2249422 "An ermes-triggered error was detected in task dir (0x396E558) for ermes number 2227915. Additional info in /usr/g/service/log/vxermesTrgLg-9102" Ran the IT diags and got the following error: 2252055 "IT-Host Card driver initialization error. This error usually means that the IT cable is disconnected or that the cable is damaged." PQR 13065908

Error message

2249422 An ermes-triggered error was detected in task dir (0x396E558) for ermes number 2227915. Additional info in /usr/g/service/log/vxermesTrgLg-9102

2218527 Scanning has stopped. Error, please try again 2218527 lx-mr scn MsaScanOpsScrn.m 916 Scanning has stopped. error please try again

2227915 The RDBM Read Header function returned the following error: 6. The IT Link is possibly disconnected or the IT boards are not communicating.

Applies to

Models
Excite, Discovery, Optima
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.