Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova10501021050512

Innova: Positioner error.

Positioner error.

Symptom

Innova: Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited, Lost All Gantry axis's drive or NO movements. Omega table elevation will not drive down. Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited, Lost All Gantry axis's drive or NO movements. Omega table elevation will not drive down. Lift failure, dose is minimized, call service

Error message

1050102 or 1050512 Lat: Pivot AXIS POT ENCODER MISMATCH DO GantryAxis Intermittent Motion stop Pivot AXIS POT ENCODER MISMATCH 1 0 - potentiometerEncoderMismatch Z:/heroic/motion control task.c 4560 1050102 or 1050512 Lat: Pivot AXIS POT ENCODER MISMATCH DO GantryAxis Intermittent Motion stop Pivot AXIS POT ENCODER MISMATCH 1 0 - potentiometerEncoderMismatch Z:/heroic/motion control task.c 4560

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.