GE Healthcare · MRSignaExcite224738618571

MCR Test Fails with The Signals Energy Level is Marginally Below the Expected Value Due to UTNS3.

Symptom

MCR Test Fails with The Signals Energy Level is Marginally Below the Expected Value See long description Diagnostic: Multi-Coil Receive II Diag Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a70 Time: 49742 Object: APS Test Task Test: 'MCR II LPCA B2, Port B' diagnostic error. The signal's energy level is marginally below the expected value. The channels are: 10, 12, 14, 16-> >

Error message

2247386 Diagnostic error. The signal's energy level is marginally below the expected value. 2247386 tempmr diagCoord rrfRxFruIso.c 1539 Diagnostic: 16 Channel Switch Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 18571 Object: APS Test Task Test: '16 Ch Switch Input Osc. AB, Port A' diagnostic error. The signal's energy level is marginally below the expected value. The channels are: 1-8->

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2366218-2 UTNS3 See Long Description 2366218-2 UTNS3 EMC ED2 Compliant .
Area
MR