GE Healthcare · MRExciteSigna2244897
Scan stops with gradient error 2244897.
Symptom
Scan stops with gradient error 2244897.
Error message
2244897 The GP is reporting a % Axis OverTemperature Fault See Long Description System will wait 5 minutes before attempting recovery. Extended Error: Check that the cabinet fan is functioning properly, rear door is closed and sealed, and that no filler panels have been removed from the front of the cabinet. Proper cooling requires a vaccum to pull air through the Gradient Amplifiers, Power Supply, and GP. If fault persists on specific PSDs write a PQR. If fault persists on many or all PSDs, replace Gradient Amplifier.
Applies to
- Models
- Excite, Signa
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Component
- SecondOpinion Control Board Assembly HFD-S-LITE 5138035-2
- Area
- MR
Full cause & step-by-step resolution