GE Healthcare · MRExciteSigna2249158
Cannot scan with TR and dynamic disable fault Dynamic disable fault Due to SSM High Voltage Supply.
Symptom
Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.
Error message
2249158 Dynamic disable fault, Front of Coil open. Dynamic Disable bias fault on front drivers. 2249158 Dynamic disable fault, Front of Coil open Dynamic Disable bias fault on front drivers. Possibly a disconnected cable between the driver board and the dynamic disable switch boxes, or an open diode in dynamic disable switch boxes on FRONT of BRM.
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR