GE Healthcare · MRExcite2224842
Table Longitudinal Drive Problems Due to Gear Reducer 40 to 1 Gear Box.
Symptom
Table Needs a Push Moving in and Cradle Unlatches When doing heavy patients the customer had to push the cradle to scan position and in some cases the cradle will come unlatched from the dog house.
Error message
2224842 SRI Hardware Error Unexpected cradle motion. Thu Jun 9 13:59:11 2005 Error: 2224842 Host: MR_PHC Process: NSP File: HPC_PORT_ID:sri Line: 99 SRI Hardware Error: Unexpected cradle motion. This error would occur if the table were moved by hand. Please run SRI diagnostics if that is not the case. Expected Cradle Velocity: -100 mm/sec Measured Cradle Velocity: 189 mm/sec
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 46-282359P2 Gear 16 mm or 40-1 speed reducer Part of the longitudinal motor drive assembly in line with motor in rear pedestal. See long description 15 46–282359P2 2 SPEED REDUCE 1 40 TO 1 NON–MAGNETIC GEAR REDUCER Note: Part Description in FEMC is simply "GEAR 16 MM"
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.