Intermittent MGD - PCI, Raven, Serr, PERR, Harrier errors that stop scanning or TPS reset failing.

Symptom

Intermittent MGD - PCI, Raven, Serr, PERR, Harrier errors that stop scanning or TPS reset failing. How to troubleshoot? see long One or more of these errors occur intermittently. How do we go about troubleshooting this type of problem? 2249410 A PCI error occured. The PCI Parity Error PERR was asserted. 2249411 A PCI error was detected, but message queue send failed 2249412 A PCI error occured. The PCI Parity Error PERR was asserted. 2249432 A PCI error occured. The PCI System Error SERR was asserted.

Error message

2249409 A PCI error was detected, but message queue send failed to the exception logging task See Long Description 2249409 lx-mr diagCoord mgd750PciErr.c 1226 Task ID: 0x6f8d18 Time: 1579446 Object: APS A PCI error was detected, but message queue send failed to the exception logging task; errno = 0x3D0002. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x1. Raven Err Address Reg: 0x2D019700; Raven Err Attribute Reg: 0x860F.

2249410 A PCI error occured. The MCP750 received a Target Abort. See Long Description 2249410 lx-mr diagCoord excLog.c 1151 Task ID: 0x3ac1d88 Time: 1579560 Object: APS A PCI error occured. The MCP750 received a Target Abort. Raven Err Enable Reg: 0x3F; Raven Err Status Reg: 0x1. Raven Err Address Reg: 0x2D019780; Raven Err Attribute Reg: 0x860F. Additional information in /usr/g/service/log/vxPciLg-1819ed

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.