Images Reconstructed but not Listed in the Browser, Only Displayed in Auto viewer.

Symptom

Images Reconstructed but not Listed in the Browser, Only Displayed in Auto viewer. See Long Desc. Caller had images in Auto Viewer but not in Browser. Caller had scanned several series since but had turned off Autoview prior to scanning the additional new series. When Autoview is turned off, it will hold onto the last images that were scanned with it turned on.

Error message

2219864 An internal error occurred in List/Select function 2219864 An internal error occured in List/Select function at Graphic Rx (9). or at 3PLANE GRx (9).

Applies to

Models
Signa, Excite, Discovery
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR