Fiber Optic Link Errors Due to IRF I/O Board.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2247160 RRF Fiber Cable Link Not Ready See Long Description Fri Jan 25 09:23:14 2008 Error: 2247160 Host: FHMR01 Process: diagCoord File: Irf.cpp Line: 272 Diagnostic: MUX BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3fa6a60 Time: 51176 Object: APS Test Task RRF FIBER CABLE LINK NOT READY The fiber link between the RRF and MGD chassis is not in a ready state after hardware initialization. This will prevent the MGD from controlling the RRF chassis. Try TPS_reset. Run RRF diagnostics.-> Extended Error: The MGD does not like the signal coming over the fiber cable connected from the IRF_IO board in the back of the MGD chassis to the DIF board in slot 10 in the front of the RRF chassis. Possible cause(s): RRF-DIF Board, GigaLink Fiber Cable, IRF I/O Board, IRF Board

2247150 The IRF driver function Exciter::ex_set_2nd_rrfdif_clk_en timed out when waiting for the RRF chassis to respond to a request. See "Long Description" 2247150 diagCoord Exciter.cpp 1398 Task ID: 0x3f1fdf8 Time: 868 Object: APS The IRF driver function Exciter::ex_set_2nd_rrfdif_clk_en timed out when waiting for the RRF chassis to respond to a request. This may happen occasionally during a scan. Device Mask = 0x1 If this 

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR5-HFO5 Open Speed with Excite Product
Component
2274196 IRF I/O Board See Long Description. Per FEMC part information, Description of part number 2274196 is an INTERFACE and REMOTE RF TRANSITION BOARD. Part Comments as of 2/8/2008 are as follows; FLS - INTERFACE & REMOTE RF TRANSITION ADDED PER RSPL. USED ON PHOENIX MGD. MSA8429 05/14/02 ROUTE TO MR FOR CONSUMABLE COLLECTION PER B. HAHLEN, 7-11-02, A. MASCH THIS PART WAS ADDED AS PART OFTHE EXCITE NPI FOR MR. 9/01/02 MS24445 BH FLS - FE ASSIST 08OCT2002 (MS2045) ADDED TO MR DISTRIBUTION LIST FOR PARTS TRACKING.
Area
MR

Full cause & step-by-step resolution