Pular para o conteúdo principal
GE Healthcare · MRExcite22471572247153224716051176224715951174

TPS Reset Fails due to RRF Chassis Digital Power Supply Unable to Download or Reset The TPS MGD or ISE.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2247157 The Master Exciter in the RRF chassis (rear, slot 7) was not detected. Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board.-> Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board. Extended Error: The master exciter provides the main clock for the RRF chassis and the IRF board. Verify that the board is powered and that its LO DETECT LED is on. If this is a 16 channel system, verify that the two fiber cables on the two DIF boards are not switched with each other. Possible cause(s): IRF Board, Exciter Board, RRF-DIF Board, GigaLink Fiber Cable

2247153 RF SUBSYSTEM NOT RESPONDING 2247153 MR1 diagCoord Irf.cpp 236 Task ID: 0x3f1fdf8 Time: 823 Object: APS RF SUBSYSTEM NOT RESPONDING The RRF chassis is not communicating to the MGD chassis. Try TPS Reset. If this problem repeats, call service. Scanning will not be possible. Verify tha

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2292265 5V RF MODULE DIGITAL POWER SUPPLY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required