2267502 Low signal during DTX1 calibration due to bent pin 9 on J5 on driver module.

Symptom

System will not scan with Error 2267502 Low Signal During DTX1 or DTX2 Calibration System will not scan with Error 2267502 Low Signal During DTX1 or DTX2 Calibration.

Error message

2267502 Low signal during DTX1 calibration See Long Description 2267502 ERROR: EM_IRF3_DTX_CAL_LOW_SIGNAL Low signal during DTX%d calibration.DTX calibration was aborted because the measured full strength signal was too low. Expected signal level: %d, Actual level: %d Extended Error: With the DTX set to full gain, the measured signal level was too low.This signal is sent from the DTX to the RF Hub which loops it back to all the RRX input channels. A low signal can be caused by a bad DTX, a bad or disconnected cable, a defect in the RF Hub, or a bad RRX.

Applies to

Models
Optima, Brivo, Excite, Signa
Product
Optima MR360 Alias; HDsv. This concept covers both Optima MR360 and Brivo MR355. Both products are produced in Beijing, China.
Software
SV20.1 Applies to Optima MR360 and Brivo MR355, a.k.a. HDsv. Both products are made in Beijing, China.
Component
SecondOpinion: Driver Module Lite 5138526
Area
MR

Full cause & step-by-step resolution