2267502 Low signal during DTX1 calibration due to bent pin 9 on J5 on driver module.
Symptom
System will not scan with Error 2267502 Low Signal During DTX1 or DTX2 Calibration System will not scan with Error 2267502 Low Signal During DTX1 or DTX2 Calibration.
Error message
2267502 Low signal during DTX1 calibration See Long Description 2267502 ERROR: EM_IRF3_DTX_CAL_LOW_SIGNAL Low signal during DTX%d calibration.DTX calibration was aborted because the measured full strength signal was too low. Expected signal level: %d, Actual level: %d Extended Error: With the DTX set to full gain, the measured signal level was too low.This signal is sent from the DTX to the RF Hub which loops it back to all the RRX input channels. A low signal can be caused by a bad DTX, a bad or disconnected cable, a defect in the RF Hub, or a bad RRX.
Applies to
- Models
- Optima, Brivo, Excite, Signa
- Product
- Optima MR360 Alias; HDsv. This concept covers both Optima MR360 and Brivo MR355. Both products are produced in Beijing, China.
- Software
- SV20.1 Applies to Optima MR360 and Brivo MR355, a.k.a. HDsv. Both products are made in Beijing, China.
- Component
- SecondOpinion: Driver Module Lite 5138526
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.