Pular para o conteúdo principal
GE Healthcare · MRExcite22650391344342016

MRIQ Zipper or Line Artifact Due to IRF I/O Board.

Symptom

MRIQ Diagonal Zipper or Line Artifact See image in long description Intermittent diagonal Zipper

Error message

2265039 APS received DRF utopia rx bus checksum error Wed Sep 8 11:47:35 2010 2265039 HDe_mfg NSP APS:DRF Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: T1flair Coil: HEAD Scan: 97/3 APS received DRF utopia rx bus checksum error. possible reason: IRF/DRF hardware error. Please perform a TPS Reset. DRF ID: 1 FE ID: 0 Int Status: 1344342016 Counter: 1->

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
2274196 IRF I/O Board See Long Description. Per FEMC part information, Description of part number 2274196 is an INTERFACE and REMOTE RF TRANSITION BOARD. Part Comments as of 2/8/2008 are as follows; FLS - INTERFACE & REMOTE RF TRANSITION ADDED PER RSPL. USED ON PHOENIX MGD. MSA8429 05/14/02 ROUTE TO MR FOR CONSUMABLE COLLECTION PER B. HAHLEN, 7-11-02, A. MASCH THIS PART WAS ADDED AS PART OFTHE EXCITE NPI FOR MR. 9/01/02 MS24445 BH FLS - FE ASSIST 08OCT2002 (MS2045) ADDED TO MR DISTRIBUTION LIST FOR PARTS TRACKING.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues