GE Healthcare · MRExcite224459222445982244578
SGA Power Supply Errors Due to SGA ACGD Power Supply.
Symptom
EPI White Pixel Test in PM Mode fails prescan with Gradient errors Everything else will run without errors, even customer EPI protocols. But EPI white pixel test will not prescan.
Error message
2244592 SGA Power Supply Under Voltage Fault 2244592 "SGA Power Supply Under Voltage Fault.\n\ Look at SGA-PS front LEDs. If BUS_OV LED lit, check PDU\n\ 420V output. If UV_X,Y,Z LED lit, swap military connectors\n\ to isolate SGA versus SGA-PS." 2244598 The GP Reported more than one %c Axis SGA Internal Wiring/Power Supply Undervoltage Faults 2244598 "The GP reported more than one %c Axis SGA Internal\n\ Wiring/Power Supply Undervoltage Faults. Refer to prior\n\ fault message. Power cycle the ACGD POWER breaker on the\n\ PDU to recover from this condition." 2244578 GP is Reporting SGA Overtemp Fault
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2317092-2 SGA ACGD PS ASM W/FRU PACKAGING FOR ACGD ONLY This SGA ACGD Power Supply can only be used on Excite (11x) or earlier systems that use the ACGD gradient subsystem. It cannot be used on Excite HD (12x) or Excite HDx (14x) systems that use the HFD or HFD-S gradient subsystem. 2317092-2 SGA ACGD PS ASM W/FRU PACKAGING FOR ACGD ONLY This SGA ACGD Power Supply can only be used on Excite (11x) or earlier systems that use the ACGD gradient subsystem. It cannot be used on Excite HD (12x) or Excite HDx (14x) systems that use the HFD or HFD-S gradient subsystem.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.