GE Healthcare · X-ray VascularInnova810020280003128000072800034

Database Inconsistency Failure

Symptom

Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited. Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. Database inconsistency failure and a Debug text of Check And Repair Requested at Browser initialization. When the customer boots the system, the image monitor displays Database inconsistency failure: Continue and call service for repair. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited.

Error message

Asynchronous IP issue error Error which error text contains "An asynchronous error has been detected by the IP"

error Code = 810020, error Text = Database inconsistency failure, debug Text += CONSISTENCY ERROR: The number of frames stored in the DataBase and on the IP are inconsistent, recovery Class = WR, major Function /= XRDataMgr System impact: Database inconsistency. Displayed message : "Database inconsistency failure: continue and call service for repair".

error Code = 2800031, error Text = Data inconsistency failure, debug Text += Check And Repair Requested upon Start Exam, recovery Class = WR, major Function /= XRSubSysMgrDL Or error code 2800007 Or error code 2800034 System impact: Database inconsistency. Displayed message : "Database inconsistency failure: continue and call service for repair".

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.