GE Healthcare · X-ray VascularInnova50111

The Gantry Pivot position does not match the In Room Display.

Symptom

The Gantry Pivot position does not match the In Room Display. When the Gantry is driven the Angle changes, but when the Gantry displays reads 0 the gantry angle indicated by the IRD is a value other than Zero. The Gantry Pivot position does not match the In Room Display. When the Gantry is driven the Angle changes, but when the Gantry displays reads 0 the gantry angle indicated by the IRD is a value other than Zero.

Error message

50111 "POS : SMA: Encoder, potentiometer failure, debug Text += pos-error=4ABC, hw-status=0001, usr-status=1000 , recovery Class = DO, major Function /= LC Gantry System impact: No gantry motion. Displayed message : None. 50111 "POS : SMA: ENCODER, POTENTIOMETER FAILURE " DO POS 0 2 HWStatus=0001 User=0000 1000 4ABC 50111 "POS : SMA: Encoder, potentiometer failure, debug Text += pos-error=4ABC, hw-status=0001, usr-status=1000 , recovery Class = DO, major Function /= LC Gantry. System impact: No gantry motion. Displayed message : None.

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.