Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova10501371050297

Lift failure, dose is minimized, call service.

Symptom

Lift failure, dose is minimized, call service. Detector lift will not move up or down. Even by using the emergency backout button. Lift failure, dose is minimized, call service. Detector lift will not move up or down. It also will not move with the emergency backout button. Lift failure, dose is not minimized, call service. Detector lift will not move up or down. It also will not move with the emergency backout button.

Error message

error Code = 1050137, error Text = LIFTBRAKE SAF FAILURE, debug Text = - Lift brake command failure. Please call service (4A73). From Validation task, recovery Class = WR, major Function /= GantryAxis System impact: No more detector lift motion. A system reset recovers. Displayed message : "!! Detector lift failure. Call Service" may be preceeded by: 1050297 DO 0 POS EVENT LOG Debug: - Emergency backout has been pressed. If so, this may be the result of the above action and not the root cause

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues