GE Healthcare · X-ray VascularInnova360201037020103602012370201236020863702086
X-Ray not available on lateral plane of a biplane system.
Symptom
X-Ray not available on lateral plane of a biplane system. After system boot, customer is moving LP to isocenter. Then X-RAY is not available on lateral plane. Following error message "Xray not available - Reset system; if problem persists, call Service" is displayed on lateral live in room monitor.
Error message
Error Code = 3602010 (FRT) or 3702010 (LAT), error Text = Generator not started. Generator switched off, debug Text = [XRTechMgr::initActor] dont allow the acquisition until the generator is powered up, recovery Class = IH, major Function /= XRTechMgr Note: May be followed by error 3602012 (3702012) AR XRTechMgr Frt (Lat): Generator communication failed Debug: Timeout during the power-up of the generator. Error Code = 3602010 (FRT) or 3702010 (LAT), error Text = Generator not started. Generator switched off, debug Text = [XRTechMgr::initActor] dont allow the acquisition until the generator is powered up, recovery Class = IH, major Function /= XRTechMgr Note: May be followed by error 3602012 (3702012) AR XRTechMgr Frt (Lat): Generator communication failed Debug: Timeout during the power-up of the generator. INX100, INX1X1: 3602086 (Frt) or 3702086 (Lat) Jedi: LVPS problem JEDI Error Code: 0xxxH INX100, INX1X1: 3602086 (Frt) or 3702086 (Lat) Jedi: LVPS problem JEDI Error Code: 0xxxH. See long description for further details on Jedi err code description. INX100, INX1X1: 3602086 (Frt) or 3702086 (Lat) Jedi: LVPS problem JEDI Error Code: 0xxxH. Standard format message is: 3602086 (3702
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.