GE Healthcare · MRExcite2224842
Carriage Has No Movement Due To PHPS.
Symptom
Cradle Will Not Move In or Out Table raised normally. LPCA will not drive in or out.
Error message
2224842 Table motion was stopped because it moved differently than expected. See Long Description 2224842 CNSJOMR2 NSP SCP:SRI 100 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Fiesta Coil: HD BodyUpper Scan: 769/9 Table motion was stopped because it moved differently than expected. This can occur if the table was moved by hand. Last Velocity Commanded: 100 mm/sec Measured Cradle Velocity: -0 mm/sec-
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5215012 MR Patient Handling Power Supply see long description 5215012 MR Patient handling power supply for assembly for MR excite 3. This is identical to the 5109368 assembly but has power supplies with a modification to allow operation upto a 200 Gauss magnetic field.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.