Unable to move the Omega Table.
Symptom
Unable to move the Omega Table. The Table top locks do not release and up/down motion does not work. The customer is unable to move the Omega Table. The Table top locks do not release and up/down motion does not work. The table is not functioning. The customer is unable to move the Omega Table. The Table top locks do not release and up/down motion does not work. errors 1050278
Error message
1050278 CPU Stepper Board Serial Communication Failure 1050278 CPU STEPPER BOARD SERIAL COM DO GantryAxis Reset From Back Plane Component(s)_Positioner_CPU_Table_Serial_communication 1 0 - Stepping board PRD no response Z:/heroic/table_geometry_task.c 1168
Applies to
- Models
- Innova
- Product
- Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- Omega table stepper board 5147596 Part #: 5147596 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: FRU New Stepper V IS6.31L FRU: Yes Status: Orderable ATTENTION: Before replacing a failed NEW STEPPER IS6.31L board (5147596): 1- do not forget to write the following LED status (ON/OFF) on the STEPPER when it is failing : +5V; +12V; +24V; RX; TX; RESET; SECURITY; READY TO MOVE. The LEDs status should be part of a reason for replacement comment e Part #: 5147596 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: FRU New Stepper V IS6.31L FRU: Yes Status: Orderable ATTENTION: Before replacing a failed NEW STEPPER IS6.31L board (5147596): 1- do not forget to write the following LED status (ON/OFF) on the STEPPER when it is failing : +5V; +12V; +24V; RX; TX; RESET; SECURITY; READY TO MOVE. The LEDs status should be part of a reason for replacement comment
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.