GE Healthcare · X-ray VascularInnova5011850211
Table pot / encoder failure is reported in the error log or Healthpage 50118 4AC3
Symptom
Table pot / encoder failure is reported in the error log or Healthpage 50118 4AC3 POS: Table Lateral pot, encoder failure reported 50118 4AC3 0x0000 HW_TBL_LONG 0x0000 HW_TBL_LONG 0x0000 HW_TBL_LONG 0x0001 HW_TBL_LAT 0x0002 HW_TBL_HEIGHT 0x0008 HW_PIB POS: Table Lateral pot, encoder failure reported 50118 or 4AC3 0x0000 HW_TBL_LONG 0x0000 HW_TBL_LONG 0x0000 HW_TBL_LONG 0x0001 HW_TBL_LAT 0x0002 HW_TBL_HEIGHT 0x0008 HW_PIB
Error message
error Code = 50118, error Text = POS Table: Encoder, potentiometer failure, debug Text += pos-error=4AC3, recovery Class = DO, major Function /= LC Gantry System impact: Degraded table motion. Displayed message : None. Status in error debug text. (0:Longitudinal, 1:Lateral, 2:Heighth, 8:PIB) error Code = 50118, error Text = POS Table: Encoder, potentiometer failure, debug Text += pos-error=4AC3, System impact: Degraded table motion. Displayed message : None. Hardware codes associated with this error: 0x0000 HW_TBL_LONG - Longitudinal Table 0x0001 HW_TBL_LAT - Lateral Table 0x0002 HW_TBL_HEIGHT -Table Height 0x0008 HW_PIB - PIB 50118 POS Table: Encoder, potentiometer failure; Debug: , HDW STATUS=0000 or HW 0001,USR STATUS=XXXX followed by 50211 POS Table: Motion control error during Bolus procedure. Both occur at the same time during a Bolus run. 50118 POS Table: Encoder, potentiometer failure; Debug: POS ERROR=4AC3, HDW STATUS=0000,USR STATUS=XXXX 50211 POS Table: Motion control error during Bolus procedure. Debug: [XRImChainPos::logPosError] Reporing the Motion Control error for GantryMotionControlError See LD for complete error. 50118 POS Table: Encoder, potentiometer failure; D
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- Omega V Table Omega V Table Omega V Table Innova 2100–IQ Case Innova 2100–IQ System can be supplied with an OMEGA IV Compact patient table, an OMEGA V long patient table, which includes a table side control with Contour Filter, or a InnovaIQ patient Table. Innova 3100/3100-IQ and 4100/4100-IQ Case Innova 3100/4100 Systems can be supplied with an OMEGA V Long patient table or InnovaIQ patient Table, which includes a table side control with Contour Filter. Illustration 2-1: Omega patient table (with TSSC) Positioner Gantry Positioner Gantry Positioner Gantry
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.