Intermittent Prescan failures with errors 2224667 and 2254345 caused by Defective Narrow Band Interface board.

Symptom

Auto and Manual Prescan Fails

Error message

2224667 Manual Prescan failed, Prescan sequencing problem, Please try again. See Long Description 2224667 NSMCGEMRAW psc ../PscMainControl.cc 1392 Exception Class: Unknown Function: Prescan: Host Manual Prescan Manual Prescan failed, Prescan sequencing problem, Please try again.

2254345 RF Amplifier Interface RF Amp serial link failure.A response packet was lost due to a UART receive error. See Long Description 2254345 SCP:../rf_nb_serial 1920 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Spin Echo Coil: BODY Scan: 50117/1 RF Amplifier Interface <-> RF Amp serial link failure. A response packet was lost due to a UART receive error. If the link is broken, a separate error message will be logged. Timer: 7134 Expired time: 40 Response bytes:

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2391146 UPM NB AMP IF BOARD, also known as ASC RF Interface Board.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.