R2 Not Updating in Manual Prescan MPS.

Symptom

R2 Not Updating in Manual Prescan MPS The numerical value in the upper-right corner of MPS always displays zero for R2 but R1 functions normally. FID is a flatline void of any noise or signal. R1's value will change when TG is increased or decreased indicating transmit power is present and R1 is making the calculation.

Error message

2236854 A data error occurred during a MEDCAM to CAP blockmode VSB data transfer. Wed Jan 12 17:18:42 2011 Error: 2236854 Host: MRS2-OC0 Process: diags File: diagsMain.m Line: 1855ISE DP: VSB Bus Test Failure.A data error occurred during a MEDCAM to CAP blockmode VSB data transfer.The data was verified by the CAP.CAP addr: 80103000HVME addr: 405b4000HData exp: 00005442HData read: 04005442HFRU(s): CERD Digital Bd, MEDCAM

2237537 A data error occurred during a CAP to MEDCAM blockmode data transfer Wed Jan 12 17:18:42 2011 Error: 2237537 Host: MRS2-OC0 Process: diags File: diagsMain.m Line: 1855ISE DP: VSB Bus Test Failure.A data error occurred during a CAP to MEDCAM blockmode data transferover the VSB bus. The data was verified by the TPS CPU.VME addr: 405b4000HData exp: 00005442H Data read: ffffffffH FRU(s): CERD Digital Board, MEDCAM

2237538 A data error occurred during a CAP to MEDCAM blockmode memory fill over Wed Jan 12 17:18:42 2011 Error: 2237538 Host: MRS2-OC0 Process: diags File: diagsMain.m Line: 1855ISE DP: VSB Bus Test Failure.A data error occurred during a CAP to MEDCAM blockmode memory fill over the VSB bus. The data was verified by the TPS CPU.VME addr: 405ac000HData

Applies to

Product
Profile 2 SPARC20 with AN8290Y gradients
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Component
2160818-7 IPGII Board SUN STYLE FRONT PANEL See Long Description IPG2 Board 2160818-7 FEMC Part Description: IPGII SUN STYLE FRONT PANEL
Area
MR

Full cause & step-by-step resolution