GE Healthcare · MRExcite224884131931
Will Not Scan With RF Errors.Power Cycle the RF Amplifier via the Circuit Breaker on the Front of the PDU And Reset TPS…
Will Not Scan With RF Errors.Power Cycle the RF Amplifier via the Circuit Breaker on the Front of the PDU And Reset TPS wqr.Excite HD.2248841 RF Amplifier Gate Fault and reports an unblank pulse width…
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2248841 RF Amplifier Gate Fault and reports an unblank pulse width exceeding spec Tue Dec 8 18:53:45 2009 Error: 2248841 Host: glrad001 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2505 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: FRFSE-XL Coil: QUADKNEE Scan: 31931/5 Analogic RF Amplifier 2 Gate Fault. Analogic RF Amplifier 2 reports an unblank pulse width exceeding spec(62-68msec). Possible causes: the PSD is exceeding the limits of the system, defective RF Interface (RFI) or defective RF Amplifier 2.-
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.