Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Missing File.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2223639 Environmental variable TPSHOST is not defined Thu Oct 23 18:03:37 2003 Error: 2223639 Host: HDC_MRI1 Process: checkForSecurityKey File: mrmailmgr.c Line: 240 Environmental variable TPSHOST is not defined.

2223640 Internal software error. Thu Oct 23 18:03:37 2003 Error: 2223640 Host: HDC_MRI1 Process: checkForSecurityKey File: mrmailpkt.c Line: 476 Internal software error.

Failed to get machine name from NSP Fri Jan 28 10:36:31 2005 Error: 0 Host: lx-mr Process: checkForSecurit yKey File: mrmailmgr.c Line: 222 Failed to get machine name from NSP