Pular para o conteúdo principal
GE Healthcare · MRExcite2249436

Can Not Boot Application Software With VRE Errors Due To VRE Configuration Problem.

Symptom

Can Not Boot Application Software With VRE Errors See Long Description 2249418 Host: LJP MR 01 Process: diagCoord File: mgdArp.c Line: 154 Task ID: 0x7effdf0 Time: 1689 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Network problem detected. Check MGD subnet network configuration. arpResolve failed for IP 10.0.0.102, numTries 2, delay 100 ticks. status = -1. Or, 2223859 Host: ghmr002 Process: NSP File: TardisReset.cc Line: 469 Exception Class: Unknown Function: Recon: Data Acquisition The TPS Reset failed because of a VRE reset failure. Or; 2223860 VRE reset failed on ICN icn1 with MAC address XX:XX:XX:XX:XX:XX Or, 2223885 Host: hde-mr Process: logermes File: vreBmcUpCheck.pl Line: 251 Exception Class: Shutdown Function: Host: NSP System startup failed due to failure to communicate with the VRE!-> Extended Error: The system startup failure is caused by the failure to communicate with all configured ICNs. Please check that power is applied to all ICNs. Please check the Ethernet connections to all ICNs. You may also run all Ethernet and IPMI tests under VRE Communication Diagnostics to verify connectivity. As a last resort, VRE Configuration may be required to recover.

Error message

2249436 Communication problem between APS and VRE. See long description Mon Jun 4 07:37:51 2007 Error: 2249436 Host: lx-mr Process: diagCoord File: cm_init.c Line: 293 Task ID: 0x6d89800 Time: 1807 Object: APS Exception Class: Unknown Function: Software/Firmware: Host Communications Communication problem between APS and VRE. The APS could not notify the VRE that communication is established.-> Extended Error: ERROR: EM_OS_CM_INIT_RPC_VRE Communication problem between APS and VRE. The APS could not notify the VRE that communication is established. Extended Error: This could be an indication of a hardware problem with the APS or VRE boards or an ethernet cabling problem on the MGD subnet. Verify the ethernet cabling on the MGD subnet, reboot VRE, and perform a TPS Reset. Run diagnostics if the problem recurs.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues