GE Healthcare · MRExcite2244590
Overcurrent Fault due to Gradient Power Supply.Will Not Scan with Gradient Errors.SecondOpinion ACGD SGA Power Supply 23…
Overcurrent Fault due to Gradient Power Supply.Will Not Scan with Gradient Errors.SecondOpinion ACGD SGA Power Supply 2334862-2 wqr.Excite.2244590 SGA-PS OverCurrent Fault.11.1
Symptom
Will Not Scan with Gradient Errors.
Error message
2244590 SGA-PS OverCurrent Fault 2244590 "SGA-PS OverCurrent Fault.\n\ Look at SGA-PS OC front LEDs to determine which SGA axis.\n\ If primary OC, replace SGA-PS. If secondary OC, swap military\n\ connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER\n\ PDU breaker to recover."
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.