Overcurrent Fault due to Gradient Power Supply.Will Not Scan with Gradient Errors.SecondOpinion ACGD SGA Power Supply 23…

Overcurrent Fault due to Gradient Power Supply.Will Not Scan with Gradient Errors.SecondOpinion ACGD SGA Power Supply 2334862-2 wqr.Excite.2244590 SGA-PS OverCurrent Fault.11.1

Symptom

Will Not Scan with Gradient Errors.

Error message

2244590 SGA-PS OverCurrent Fault 2244590 "SGA-PS OverCurrent Fault.\n\ Look at SGA-PS OC front LEDs to determine which SGA axis.\n\ If primary OC, replace SGA-PS. If secondary OC, swap military\n\ connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER\n\ PDU breaker to recover."

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.