Pular para o conteúdo principal
GE Healthcare · MRExciteDiscoveryOptima22471442265076

CWA Application Software does not complete the boot process with TPS Download failure.

Symptom

Application Software does not complete the boot process with TPS Download failure

Error message

2247144 The IRF driver function Dcerddif::rifdif_get_dcerd_variation was called with a bad parameter value. Bad parm = 0 This is a software error.-> Error: 2247144 The IRF driver function Dcerddif::rifdif_get_dcerd_variation was called with a bad parameter value. Bad parm = 0 This is a software error.->

2265076 Old DCERD Digital Board Found Error: 2265076 Proc: diagCoord File: Dcerd.cpp Line: 258Object: APSException class: unkownFunction: Acquisition Pulse Generation: AcquisitionOLD DCERD DIGITAL BOARD FOUNDThe old revision of DCERD Digital Board is found in the DCERD chassis. Replace to newer version.

Applies to

Models
Excite, Discovery, Optima
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Area
MR

Full cause & step-by-step resolution