GE Healthcare · MRSignaExcite222554222255262225540
2225540 Dynamic Disable Bias Fault on Front Drivers Due to the SSM.
Symptom
Unable to Scan with Inductive or Direct Drive Bias Fault Unable to Scan with Inductive or Direct Drive Bias Fault. Sat Apr 23 10:21:22 2005 Error: 2225526 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6100 Inductive or Direct Drive fault, or Body TR switch fault due to impedance mismatch, path short Drive bias fault. Possibly a shorted diode in the driver switches, or a short circuit load.
Error message
2225542 Dynamic Disable bias fault on back drivers Dynamic Disable bias fault on back drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on back of BRM. 2225526 Inductive Or Direct Drive Fault, Or Body TR Switch Fault Due To Impedance Mismatch, Path Short Drive Bias Fault 2225526 INDUCTIVE OR DIRECT DRIVE FAULT, OR BODY TR SWITCH FAULT DUE TO IMPEDANCE MISMATCH, PATH SHORT DRIVE BIAS FAULT. POSSIBLY A SHORTED DIODE IN THE DRIVER SWITCHES, OR A SHORT CIRCUIT LOAD 2225540 Dynamic disable fault, Front of Coil short. Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM. 2225540 Dynamic disable fault, Front of Coil short Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM.
Applies to
- Models
- Signa, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.