Table Longitudinal Drive Problems Due to Gear Reducer 40 to 1 Gear Box.

Symptom

Table is Getting Stuck and Cannot Move In or Out With Patient Table Is Getting Stuck And Cannot Move In Or Out With Patient.

Error message

2224842 Table motion was stopped because it moved differently than expected. See Long Description 2224842 CNSJOMR2 NSP SCP:SRI 100 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Fiesta Coil: HD BodyUpper Scan: 769/9 Table motion was stopped because it moved differently than expected. This can occur if the table was moved by hand. Last Velocity Commanded: 100 mm/sec Measured Cradle Velocity: -0 mm/sec-

2224864 Table motion was stopped because it was moving faster than expected. Tue Jun 2 10:59:30 2009 Error: 2224864Host: IMCMROP02 Process: NSP File: SCP:SRI Line: 99Exception Class: Unknown Severity: UnknownPSD: Gradient Echo Coil: 8CARDIAC Scan: 3806/13Function: Patient Handling: Longitudinal Motion. Table motion was stopped because it was moving faster than expected. This can occur if the table was moved by hand. Limit: 150 mm/sec, Last Velocity Commanded: -100 mm/secMeasured Cradle Velocity: -169 mm/sec->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
46-282359P2 Gear 16 mm or 40-1 speed reducer Part of the longitudinal motor drive assembly in line with motor in rear pedestal. See long description 15 46–282359P2 2 SPEED REDUCE 1 40 TO 1 NON–MAGNETIC GEAR REDUCER Note: Part Description in FEMC is simply "GEAR 16 MM"
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.