GE Healthcare · MRExcite22248422224864
Table Longitudinal Drive Problems Due to Gear Reducer 40 to 1 Gear Box.
Symptom
Table is Getting Stuck and Cannot Move In or Out With Patient Table Is Getting Stuck And Cannot Move In Or Out With Patient.
Error message
2224842 Table motion was stopped because it moved differently than expected. See Long Description 2224842 CNSJOMR2 NSP SCP:SRI 100 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Fiesta Coil: HD BodyUpper Scan: 769/9 Table motion was stopped because it moved differently than expected. This can occur if the table was moved by hand. Last Velocity Commanded: 100 mm/sec Measured Cradle Velocity: -0 mm/sec- 2224864 Table motion was stopped because it was moving faster than expected. Tue Jun 2 10:59:30 2009 Error: 2224864Host: IMCMROP02 Process: NSP File: SCP:SRI Line: 99Exception Class: Unknown Severity: UnknownPSD: Gradient Echo Coil: 8CARDIAC Scan: 3806/13Function: Patient Handling: Longitudinal Motion. Table motion was stopped because it was moving faster than expected. This can occur if the table was moved by hand. Limit: 150 mm/sec, Last Velocity Commanded: -100 mm/secMeasured Cradle Velocity: -169 mm/sec->
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 46-282359P2 Gear 16 mm or 40-1 speed reducer Part of the longitudinal motor drive assembly in line with motor in rear pedestal. See long description 15 46–282359P2 2 SPEED REDUCE 1 40 TO 1 NON–MAGNETIC GEAR REDUCER Note: Part Description in FEMC is simply "GEAR 16 MM"
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.