GE Healthcare · MRExciteSigna22488142248816
Power monitor fault due to SSM.Unable to scan with Power Monitor Faults.SecondOpinion System Support Module 2160200-43 w…
Power monitor fault due to SSM.Unable to scan with Power Monitor Faults.SecondOpinion System Support Module 2160200-43 wqr.Excite.2248814 Power Monitor 1 Fault.11.1
Symptom
Unable to scan with Power Monitor Faults
Error message
2248814 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. This could also occur when power cycling SSM. 2248816 RF Interface (RFI) High Voltage Relay Fault 2248816, RF Interface (RFI) High Voltage Relay Fault. Possible causes: RF Monitor Cable to System Support Module (SSM) is disconnected or defective.
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.